TY - JOUR
T1 - Pacing lead implantation without live fluoroscopy
T2 - Feasibility of acute success in the live canine model
AU - Choudhuri, Indrajit
AU - Krum, David
AU - Hare, John
AU - Kelley, Jim
AU - Becker, David
AU - Mortada, M. Eyman
AU - Akhtar, Masood
AU - Sra, Jasbir
PY - 2009/8
Y1 - 2009/8
N2 - Pacing Lead Placement Without Live Fluoroscopy. Introduction: Fluoroscopic visualization for transvenous pacing lead placement necessitates lead shielding to minimize radiation exposure. An electromagnetic (EM) navigation system that integrates real-time intracardiac tracking within an anatomic navigation environment may provide an effective alternative for lead delivery that obviates live fluoroscopy. We assessed feasibility of pacing lead implantation with electromagnetic tracking guided solely by radiographic virtual navigation and compared this to fluoroscopy-guided implants in a canine model. Methods: Seven mongrel dogs with normal hearts were randomized to 47 pacing lead placements in the right atrium (RA) or right ventricle (RV) guided by single-plane fluoroscopy, or an experimental EM navigation system guided by registered fluoroscopic snapshots obtained before implant (EMN). Ability to achieve successful lead delivery acutely was assessed, and pacing parameters as well as fluoroscopy and implant times were measured. Means were compared using a paired t-test. Results: All lead delivery attempts were acutely successful. One atrial lead dislodged with EMN, resulting in 46 successful pacing attempts. There was no statistical difference in pacing parameters and time for lead placement between the approaches (EMN vs fluoroscopic navigation [mean ± SD]: RA threshold 1.15 V ± 0.98 V vs 1.95 V ± 0.98 V [P = NS], RV threshold 1.18 V ± 0.58 V vs 1.42 V ± 0.63 V [P = NS], implant time 4:38 ± 2:37 minutes vs 4:44 ± 2:38 minutes [P = NS]). No live fluoroscopy was required for EMN implants. Conclusion: Pacing lead placement with an EM system guided by preprocedural fluoroscopic views is feasible and comparable to fluoroscopic navigation, and avoids the use of live fluoroscopy. (J Cardiovasc Electrophysiol, Vol. 20, pp. 916-922, August 2009)
AB - Pacing Lead Placement Without Live Fluoroscopy. Introduction: Fluoroscopic visualization for transvenous pacing lead placement necessitates lead shielding to minimize radiation exposure. An electromagnetic (EM) navigation system that integrates real-time intracardiac tracking within an anatomic navigation environment may provide an effective alternative for lead delivery that obviates live fluoroscopy. We assessed feasibility of pacing lead implantation with electromagnetic tracking guided solely by radiographic virtual navigation and compared this to fluoroscopy-guided implants in a canine model. Methods: Seven mongrel dogs with normal hearts were randomized to 47 pacing lead placements in the right atrium (RA) or right ventricle (RV) guided by single-plane fluoroscopy, or an experimental EM navigation system guided by registered fluoroscopic snapshots obtained before implant (EMN). Ability to achieve successful lead delivery acutely was assessed, and pacing parameters as well as fluoroscopy and implant times were measured. Means were compared using a paired t-test. Results: All lead delivery attempts were acutely successful. One atrial lead dislodged with EMN, resulting in 46 successful pacing attempts. There was no statistical difference in pacing parameters and time for lead placement between the approaches (EMN vs fluoroscopic navigation [mean ± SD]: RA threshold 1.15 V ± 0.98 V vs 1.95 V ± 0.98 V [P = NS], RV threshold 1.18 V ± 0.58 V vs 1.42 V ± 0.63 V [P = NS], implant time 4:38 ± 2:37 minutes vs 4:44 ± 2:38 minutes [P = NS]). No live fluoroscopy was required for EMN implants. Conclusion: Pacing lead placement with an EM system guided by preprocedural fluoroscopic views is feasible and comparable to fluoroscopic navigation, and avoids the use of live fluoroscopy. (J Cardiovasc Electrophysiol, Vol. 20, pp. 916-922, August 2009)
KW - Cardiac pacing
KW - Electromagnetic navigation
KW - Image guidance
KW - Radiation
KW - Virtual fluoroscopy
UR - https://www.scopus.com/pages/publications/68149109937
UR - https://www.scopus.com/inward/citedby.url?scp=68149109937&partnerID=8YFLogxK
U2 - 10.1111/j.1540-8167.2009.01451.x
DO - 10.1111/j.1540-8167.2009.01451.x
M3 - Article
C2 - 19298557
AN - SCOPUS:68149109937
SN - 1045-3873
VL - 20
SP - 916
EP - 922
JO - Journal of Cardiovascular Electrophysiology
JF - Journal of Cardiovascular Electrophysiology
IS - 8
ER -